MD Scientific
MD Scientific Logo
Specialister i kromatografi
   
MD Scientific
MD Scientific
MD Scientific
     
GPC/SEC systemer
GPC/SEC detectorer & software
DSV systemer
HPLC detektionssystemer
UV imaging detektor
Osmometre
Sievers NOA 280i
Gasgeneratorer
Peptide synthesizers
Kolonner
Standarder & kits
Service & kurser
Kontakt
Om os
 

Troubleshooting Guide — Liquid Sampling - NOA 280i


A list of possible problems with the chemiluminescence detector appears below. If the problem is not listed below or assistance is needed, contact us - see contact info below..

Low Sensitivity
 Possible Causes
Action Required
Dirty reaction chamber Clean reaction chamber
Clogged frit restrictor Clean or replace the frit restrictor
Leaks in the purge system For VCl3 reagent, check if reagent vacuum distills, if not confirm Teflon sleeve on NaOH trap is tight.
Replace septum
Confirm Swagelok connections are tight, ferrules are present and no glass is present in fitting.
Confirm connections to filter are tight.
Purge gas flow too high or too low Record cell pressure when frit restrictor is open to atmosphere (filter line disconnected), then adjust needle valve to match cell pressure when purge system is connected to frit restrictor.
Clogged hydrophobic filter Clean or replace filter
Water bath temperature too low (VCl3 reagent) Set water bath to 95° C
 
Poor Repeatability
Possible Causes
Action Required
Leaks in purge system See above
Syringe contaminated or leaking Clean syringe and needle (should get no peak with empty syringe injection)
Replace syringe
Sample not injected into reagent (for µL volumes) Adjust volume of reagent so that end of needle is below reagent levels
 
VCl3 Reagent Vacuum Distills
Possible Causes
Action Required
Purge flow rate too low Adjust flow rate to match cell pressure when frit restrictor is open to atmosphere
 
High Baseline or Rising Baseline
Possible Causes Action Required
Leaks in purge system See above
Dirty Purge Vessel Clean purge vessel with laboratory detergent
Injection of material slowly converted to NO (e.g., L-NAME or other nitro compounds when using the VCl3 reagent) Nitro-containing inhibitors (L-NAME, nitro-arginine, 7-nitroindazole) cannot be used with VCl3 reagent.
Replace reagent.
 
Peaks observed without injections
Possible Causes Action Required
Leaks in purge system See above and also check for leaks in tubing from purge gas regulator to purge vessel
Contamination in reagent Will sometimes see peaks with VCl3 reagent due to contamination; allow the reagent to purge for 10-15 minutes.

 

NOA 280i (startpage)

For flere detaljer og pris, kontakt venligst MD Scientific via e-mail: info@md-scientific.dk eller telefon: 7027 8565.

Top

 

MD Scientific
MD Scientific
MD Scientific
MD Scientific ApS • Denmark • tlf. 7027 8565 • fax 7027 8566 • info@md-scientific.dk