The ActiPix™ D100 UV Area Imaging System
The Paraytec ActiPix™ D100 is a miniature capillary UV detector, designed for on-line monitoring at a single UV wavelength, particularly for use with the separation methods of liquid chromatography and capillary electrophoresis.
The ActiPix™ D100 UV area imaging system incorporates Paraytec's patented 'self referencing' detection technology. This product is transforming UV detection with its high sensitivity, nanolitre sample requirements and wide dynamic range. Area imaging allows real time multiplex analysis as well as unlocking completely new applications for UV detection.
|
|
Key Product Features - The ActiPix™
- Miniature UV detector that can be coupled with existing lab hardware, e.g. syringe pumps, capillary electrophoresis (CE) systems, nano- and capillary-LC (liquid chromatography) with or without a MS (mass spectrometer) system
- Capillary-based detection, in-line or discrete, allowing sub-microlitre sample determination.
- High spatial resolution without sacrificing sensitivity.
- ActiPix™ D series available as single or multiplexed configurations - up to 8 capillaries in one detector.
- Five orders of magnitude linear dynamic range.
- Powerful active pixel sensor provides real-time images.
- Can be used down to 190 nm in the UV.
Description
The ActiPix™ D100 consists of a control box and a small remote sensor head containing a capillary cartridge. The remote head is designed for positioning exactly at your required detection point, and can be literally clipped in to an analysis stream. Its small size (6.0 x 6.5 x 11cm) means that it is ideal for in-line detection with LC, nano-LC, LC-MS and CE.
The miniature size allows the detector to be used as a ‘clip-on’ accessory linked to existing separations instrumentation and in line with a mass spectrometer. Detection is performed at a selected wavelength by means of exchangeable filters. Light is delivered via a fibre optic coupled light source. Readout of signals on the active pixel sensor and absorbance calculations are carried out at MHz rates, with all processing done in hardware on custom electronic assemblies. Processed data is output in real-time to a computer using an industry-standard high-speed USB or Ethernet cable.
For a more detailed explanation of the theory behind the ActiPix™ Nano-Sizing system, go to our downloads section for the following publications:
Specifications
- Light source: Pulsed xenon lamp
- Fibre optic cable: Round-to-line (1 m in total)
- Wavelengths: 190 - 1200 nM, filter selectable
- Filter Bandpass: 10 nm
- Peptide limit of quantification: 100 ng/mL
- Baseline noise in absorbance: 5 µAU RMS (1 second time constant)
- Linear dynamic range: 5 orders of magnitude: 0.005 mAU – 500 mAU
- Spatial resolution: 70 µm
- Pixel size: 7 µm x 7 µm
- Imager area: 9 x 7 mm (L x W)
- Sensor head size: 6 x 6.5 x 11 cm (H x W x L)
- Control box size: 14 x 22 x 25 (H x W x L)
- Total system weight: 7.5 kg
- Power: 100-240 V AC 50/60 Hz
- Environment: 10 - 30 ° C
- Safety: Certified to UL61010-1:2004, IEC61010-1:2001
MD Scientific præsenterede den nye, prisbelønnede UV imaging detektor ActiPix D100 på Scanlab: Se pressemeddelelse.
Paraytec's ActiPix™ D100 has been awarded the Pittcon 2007 Editors' Choice Silver Award for Best New Product at the Pittsburgh Conference on Analytical Chemistry and Applied Spectroscopy (Pittcon), USA and The R &D Magazines' R&D 100 prize for 2007.
 |
|
|
|
ActiPix D100 UV imaging detector |
|
Paraytec's ActiPix D100, showing the relocatable remote detector head and benchtop control box |
| |
|
|
| |
|
|
|
|
|