The Viscotek TDAmax is a complete, advanced, multidetector GPC/SEC system suitable for all macromolecular applications.
|
|
OmniSEC |
 |
| Acquisition |
|
6 channels, 5Hz, unlimited acquisition time |
 |
| Calculation modules |
|
Conventional, Universal, Light Scattering, Triple, Copolymer, Protein specific, FIPA |
 |
| Results output |
|
Mn, Mw, Mz, Mp, Mw/Mn, IV [η], Rh, Rg, Branching Number, Branching Frequency, Number of Arms, Weight Fraction, Concentration, MH-a, MH-k, dn/dc, dA/dc. A2 |
 |
| Minimum PC specification |
|
Windows XP or higher, 100 GB HD, 2GB RAM, 1.8 GHz processor,
Graphics 1024x768, 16-bit colour depth |
 |
|
|
TDA |
 |
| Temperature control |
|
Ambient to 80ºC all detector cells and columns |
 |
| Detector configuration |
|
All detector cells in series for full signals |
 |
| Column capacity |
|
Up to 5 x 30cm GPC/SEC columns |
 |
| Interface |
|
6 channel digital interface to OmniSEC software
2 analog inputs, RI and UV/LS |
 |
| Dimensions |
|
41cm x 28cm x 54cm (W x H x D); 30Kg |
 |
|
|
Light scattering detector |
 |
| Measuring principle |
|
RALS, 90º angle; LALS, 7º angle, with high efficiency optics |
 |
| Light Source |
|
Temperature controlled laser diode, 3mW, 670nm |
 |
| Cell |
|
Ultra low volume 18μL |
 |
| Data rate |
|
100Hz collection, DSP with 5Hz to OmniSEC |
 |
|
|
Refractive index detector |
 |
| Measuring principle |
|
Deflection |
 |
| Light Source |
|
LED |
 |
| Cell |
|
45° quartz glass, 12μL volume |
 |
| Purging |
|
3 modes, Manual, Automatic and Programmable |
 |
| Data rate |
|
100Hz collection, DSP with 5Hz to OmniSEC |
 |
|
|
Viscometer detector |
 |
| Measuring principle |
|
Inert, 4 capillary differential Wheatstone bridge configuration |
 |
| Transducers |
|
Digital Inert Transducer technology with over-pressure protection |
 |
| Measuring volume |
|
18μL |
 |
| Purging |
|
3 modes, Manual, Automatic and Programmable |
 |
| Data rate |
|
100Hz collection, DSP with 5Hz to OmniSEC |
 |
|
|
UV detector |
 |
| Wavelength |
|
190 to 740nm |
 |
| Light Source |
|
Deuterium; Tungsten halogen is optional |
 |
| Cell volume |
|
10μL |
 |
| Data rate |
|
Analog collection of single wavelength (programmable), A-to-D with 5Hz to OmniSEC |
 |
|
|
UV-PDA detector |
 |
| Wavelength/resolution |
|
190 to 500nm or 430 to 710nm on 256 diode array |
 |
| Light Source |
|
Deuterium (190-500nm) or Tungsten (430-710nm) |
 |
| Cell volume |
|
10μL |
 |
| Data rate |
|
256 channels at 1Hz direct to OmniSEC |
 |
|
|
GPCmax |
 |
| Operation and Control |
|
By OmniSEC software or from front panel |
 |
| Dimensions |
|
55cm x 38cm x 53cm (W x H x D); 37kg |
 |
|
|
Degasser |
 |
| Channels/volume |
|
Two channels, 8ml per channel |
 |
| Performance |
|
<0.5 ppm oxygen at 0.5 mL/min |
 |
| Eluent Sensor |
|
User selectable, stops GPC pump when eluent runs out |
 |
|
|
Pump |
 |
| Flow rate range |
|
0.01 to 9.99 mL/min |
 |
| Pulsation |
|
Less than 1% (measured with viscometer) |
 |
| Pressure Reading |
|
MPa or PSI |
 |
| Soft start/stop |
|
User programmable in mL/min/min |
 |
|
|
Autosampler |
 |
| Capacity (Standard) |
|
120 vials, freely programmable position sequence |
 |
| Range |
|
Variable injection volume, between 20 and 150μL |
 |
| Reproducibility |
|
Better than 0.5 % |
 |
| Carryover |
|
Less than 0.1%, depending on wash program |
 |
| Cooling option |
|
4°C to 40°C |
 |
| Capacity (with cooling option) |
|
60 vials cooled, freely programmable position sequence |
 |
| Heated tray option |
|
65°C fixed in one tray |
 |
| Capacity (with heating option) |
|
60 vials heated, 60 unheated, freely programmable position sequence |
 |
|